You are here: Experiment Guide > Imaging Modes > TappingMode AFM > Engage and Scan

Engage and Scan

Engage

  1. Select Microscope > Engage or click the Engage icon on the toolbar. A pre-engage check begins, followed by Z-stage motor motion.
  1. After engaging, it may be necessary to recheck the Cantilever Tune parameters:
    1. Select Microscope > Cantilever Tune or click the Tune icon in the Workflow Toolbar.
    2. In the Tip offset box enter a value between 50 and 200 nm and click OK.
    3. Adjust the Drive frequency if necessary.

Optimize Scan Parameters

 
  1. In the Scan panel, set Slow Scan Axis to Disabled. This function is available in the Expanded Mode.
 
  1. Observe the agreement between the Trace and Retrace lines. Adjust the Scan Rate, Gains, and Amplitude Setpoint to bring these two lines in to coincidence. See Parameter Settings for a more detailed discussion of these inputs.

Withdraw the Tip

  1. Select RealTime > Withdraw or click the Withdraw icon in the Workflow Toolbar. The SPM stops scanning, and ascends to the sample clearance height defined in the SPM Parameters menu.
 
  1. Replace or move the sample.
 
  1. Use the Focus Surface command to move the SPM up if you desire more clearance between the tip and sample.

CAUTION: Never withdraw samples without verifying that the tip has adequate clearance during the entire sample removal sequence.

Previous steps:

  1. Set Up the Experiment
  2. Scanner, Probe, and Sample Preparation
  3. Locate the Probe Tip
  4. Tune the Cantilever
  5. Focus Surface
  6. Check Initial Scan Parameters

Related Topics

 

www.bruker.com Bruker Corporation
www.brukerafmprobes.com 112 Robin Hill Rd.
nanoscaleworld.bruker-axs.com/nanoscaleworld/ Santa Barbara, CA 93117
   
  Customer Support: (800) 873-9750
  Copyright 2010, 2011. All Rights Reserved.